GL7

300mK sorption cooler

Gram per gram, Helium 3 is the world’s most expensive substance. CRC’s gas-light sub-Kelvin sorption coolers have been designed to make the best possible use of this scarce resource. By buffering the ultra-cold stage with either one or two intermediate stages, we reduce the heat load on the ultra-cold head and hence lower the running temperature and lengthen the run time. As our coolers are sealed systems, no helium gas is consumed during operation and our units never need to be recharged or re-filled. Our latest design, the 300mK continuous cooler, even offers unlimited run time.

Downloads available:

Contact Us

View Gallery

Standard GL7 — 16L/2L

Overview

The GL7 is a two-stage, single-shot sorption cooler that is designed to interface to a low-power 4K precooler, such as a Gifford-McMahon (GM) cryocooler. It consists of a 1K 4He pre-cooler and a 300mK 3He module. A standard 16L/2L (4He/3He) GL7 will run for 12 hours under 30µW of thermal load on the 3He cold head at a temperature ≤0.33K. This system can be customised to operate with higher loads and have increased run time.

The GL7 has no external gas handling or pumped gas circulation. The sorption cooler only requires electrical connections to the outside of the cryostat. A thermometry box is needed to measure the sensors along with a programable power supply to control the heaters.

Typical Load Response

Standard GL7 — 16L/2L — Typical Load Response. Includes the supplied labels, units and annotations. Enlarge figure
Standard GL7 — 16L/2L — Typical Load Response

Typical run

Standard GL7 — 16L/2L — Typical run. Includes the supplied labels, units and annotations. Enlarge figure
Standard GL7 — 16L/2L — Typical run

Typical cycle

Standard GL7 — 16L/2L — Typical cycle. Includes the supplied labels, units and annotations. Enlarge figure
Standard GL7 — 16L/2L — Typical cycle

GL7 Pin-Out

GL7 pinouts to 25-pin micro-D SSP
Function Type PIN # Drive current or voltage Resistance
3He HEAD THERMOMETER V+
3He HEAD THERMOMETER V-
3He HEAD THERMOMETER I+
3He HEAD THERMOMETER I-
Resistance thermometer 1
14
2
15
100nA AC
Or low voltage Driver e.g. V<0.5mV
1kΩ
n/c 3
4He HEAD THERMOMETER V+
4He HEAD THERMOMETER V-
Resistance thermometer 4
16
1µA Constant Current 1kΩ
DIODE FILM BURNER I+
DIODE FILM BURNER I-
Diode thermometer 5
17
10 µA
DIODE 4He PUMP I+
DIODE 4He PUMP I-
Diode thermometer 6
18
10 µA
DIODE 3He PUMP I+
DIODE 3He PUMP I-
Diode thermometer 7
19
10 µA
DIODE 4He SWITCH I+
DIODE 4He SWITCH I-
Diode thermometer 8
20
10 µA
DIODE 3He SWITCH I+
DIODE 3He SWITCH I-
Diode thermometer 9
21
10 µA
HEATER 4He SWITCH I+
HEATER 4He SWITCH I-
Low power heater (a few mW) 10
22
4 to 5 Volts 10kΩ
HEATER 3He SWITCH I+
HEATER 3He SWITCH I-
Low power heater (a few mW) 11
23
4 to 5 Volts 10kΩ
HEATER 4He PUMP I+
HEATER 4He PUMP I-
High power heater (up to about 2W) 12
24
50 to 100 mA
15 to 30 V
300Ω
HEATER 3He PUMP I+
HEATER 3He PUMP I-
High power heater (up to about 2W) 13
25
51 to 100 mA
15 to 30 V
300Ω

Connector diagrams

Standard GL7 — 16L/2L — Connector diagrams. Includes the supplied labels, units and annotations. Enlarge figure
Standard GL7 — 16L/2L — Connector diagrams

Assembly diagram

Standard GL7 — 16L/2L — Assembly diagram. Includes the supplied labels, units and annotations. Enlarge figure
Standard GL7 — 16L/2L — Assembly diagram

Large GL7 — 20L/5L

Overview

The GL7 is a two-stage, single-shot sorption cooler that is designed to interface to a low-power 4K precooler, such as a Gifford-McMahon (GM) cryocooler. It consists of a 1K 4He pre-cooler and a 300mK 3He module. A ‘large’ 20L/5L (4He/3He) GL7 will run for 24 hours under 50µW of thermal load on the 3He cold head at a temperature ≤0.34K. This system can be customised to operate with higher loads and have increased run time.

The GL7 has no external gas handling or pumped gas circulation. The sorption cooler only requires electrical connections to the outside of the cryostat. A thermometry box is needed to measure the sensors along with a programable power supply to control the heaters.

Typical Load Response

Large GL7 — 20L/5L — Typical Load Response. Includes the supplied labels, units and annotations. Enlarge figure
Large GL7 — 20L/5L — Typical Load Response

Typical run

Large GL7 — 20L/5L — Typical run. Includes the supplied labels, units and annotations. Enlarge figure
Large GL7 — 20L/5L — Typical run

Typical cycle

Large GL7 — 20L/5L — Typical cycle. Includes the supplied labels, units and annotations. Enlarge figure
Large GL7 — 20L/5L — Typical cycle

GL7 Pin-Out

GL7 pinouts to 25-pin micro-D SSP
Function Type PIN # Drive current or voltage Resistance
3He HEAD THERMOMETER V+
3He HEAD THERMOMETER V-
3He HEAD THERMOMETER I+
3He HEAD THERMOMETER I-
Resistance thermometer 1
14
2
15
100nA AC
Or low voltage Driver e.g. V<0.5mV
1kΩ
n/c 3
4He HEAD THERMOMETER V+
4He HEAD THERMOMETER V-
Resistance thermometer 4
16
1mA Constant Current 1kΩ
DIODE FILM BURNER I+
DIODE FILM BURNER I-
Diode thermometer 5
17
10 mA
DIODE 4He PUMP I+
DIODE 4He PUMP I-
Diode thermometer 6
18
10 mA
DIODE 3He PUMP I+
DIODE 3He PUMP I-
Diode thermometer 7
19
10 mA
DIODE 4He SWITCH I+
DIODE 4He SWITCH I-
Diode thermometer 8
20
10 mA
DIODE 3He SWITCH I+
DIODE 3He SWITCH I-
Diode thermometer 9
21
10 mA
HEATER 4He SWITCH I+
HEATER 4He SWITCH I-
Low power heater (a few mW) 10
22
4 to 5 Volts 10kΩ
HEATER 3He SWITCH I+
HEATER 3He SWITCH I-
Low power heater (a few mW) 11
23
4 to 5 Volts 10kΩ
HEATER 4He PUMP I+
HEATER 4He PUMP I-
High power heater (up to about 2W) 12
24
50 to 100 mA
15 to 30 V
300Ω
HEATER 3He PUMP I+
HEATER 3He PUMP I-
High power heater (up to about 2W) 13
25
51 to 100 mA
15 to 30 V
300Ω

Connector diagrams

Large GL7 — 20L/5L — Connector diagrams. Includes the supplied labels, units and annotations. Enlarge figure
Large GL7 — 20L/5L — Connector diagrams

Assembly diagram

Large GL7 — 20L/5L — Assembly diagram. Includes the supplied labels, units and annotations. Enlarge figure
Large GL7 — 20L/5L — Assembly diagram

Quick Enquiry